Emerson 4500 Benutzerhandbuch

Seite von 70
Reference Manual 
00809-0100-4027, Rev AA
December 2005
3-21
Rosemount 4500
4. Select the back arrow from the HOME/ONLINE screen. The HART 
Communicator menu appears.
5. Select 1: Offline, 2: Saved Configuration, 1: Module Contents to reach 
the MODULE CONTENTS menu.
6. Use the DOWN ARROW to scroll through the list of configurations in the 
memory module, and use the RIGHT ARROW to select and retrieve the 
required configuration.
7. Select 1: Edit.
8. Select 1: Mark All.
9. Select F2 SAVE.
10. Use the DOWN ARROW to scroll through the list of configurations in the 
memory module, and use the RIGHT ARROW to select the configuration 
again.
11. Select 3: “Send” to download the configuration to the transmitter.
12. Select OK after the control loop is set to manual.
13. After the configuration has been sent, select OK to acknowledge that the 
loop can be returned to automatic control.
When finished, the HART Communicator informs you of the status. Repeat 
Steps 3 through 13 to configure another transmitter.
NOTE
The transmitter receiving cloned data must have the same software version 
(or later) as the original transmitter.
AMS creating a Reusable Copy
To create a reusable copy of a configuration perform the following procedure: 
1. Select View then User Configuration View from the menu bar (or click the 
toolbar button).
2. In the User Configuration window, right click and select New from the 
context menu. 
3. In the New window, select a device from the list of templates shown, and 
click OK.
4. The template is copied into the User Configurations window, with the tag 
name highlighted; rename it as appropriate and press Enter.
NOTE
A device icon can also be copied by dragging and dropping a device template 
or any other device icon from AMS Explorer or Device Connection View into 
the User Configurations window. 
The “Compare Configurations” window appears, showing the Current values 
of the copied device on one side and mostly blank fields on the other (User 
Configuration) side.
5. Transfer values from the current configuration to the user configuration 
as appropriate or enter values by typing them into the available fields.
6. Click Apply to apply the values, or click OK to apply the values and close 
the window.