Owner’s Record
The model and serial numbers are located at the rear. 
Record these numbers in the space provided below.
Refer to these numbers whenever you call upon your Sony 
dealer regarding this product.
Model No. ____________________
Serial No. ____________________
To reduce the risk of fire or electric shock, do not 
expose this apparatus to rain or moisture.
To avoid electrical shock, do not open the 
cabinet. Refer servicing to qualified personnel 
only.
THIS APPARATUS MUST BE EARTHED.
To disconnect the main power, unplug the AC IN 
connector.
Warning
This unit has no power switch.
When installing the unit, incorporate a readily accessible 
disconnect device in the fixed wiring, or connect the power 
plug to an easily accessible socket-outlet near the unit.
If a fault should occur during operation of the unit, operate 
the disconnect device to switch the power supply off, or 
disconnect the power plug.
For the customers in the U.S.A.
This equipment has been tested and found to comply with 
the limits for a Class A digital device, pursuant to Part 15 
of the FCC Rules. These limits are designed to provide 
reasonable protection against harmful interference when 
the equipment is operated in a commercial environment. 
This equipment generates, uses, and can radiate radio 
frequency energy and, if not installed and used in 
accordance with the instruction manual, may cause 
harmful interference to radio communications. Operation 
of this equipment in a residential area is likely to cause 
harmful interference in which case the user will be 
required to correct the interference at his own expense. 
You are cautioned that any changes or modifications not 
expressly approved in this manual could void your 
authority to operate this equipment.
All interface cables used to connect peripherals must be 
shielded in order to comply with the limits for a digital 
device pursuant to Subpart B of Part 15 of FCC Rules.
For the customers in Canada
This unit has been certified according to Standard CSA 
C22.2 No.601.1.
For the customers in the U.S.A and Canada
When you use this product connected to 240 V single 
phase, be sure to connect this product to a center tapped 
circuit.
Important safeguards/notices for use in the 
medical environments
1. All the equipments connected to this unit shall be 
certified according to Standard IEC60601-1, 
IEC60950-1, IEC60065 or other IEC/ISO Standards 
applicable to the equipments.
2. Furthermore all configurations shall comply with the 
system standard IEC60601-1-1. Everybody who 
connects additional equipment to the signal input part 
or signal output part configures a medical system, and 
is therefore, responsible that the system complies with 
the requirements of the system standard IEC60601-1-1. 
If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected to 
other equipment.
4. For this particular equipment, all accessory equipment 
connected as noted above, must be connected to mains 
via an additional isolation transformer conforming with 
the construction requirements of IEC60601-1 and 
providing at least Basic Insulation.
5. This equipment generates, uses, and can radiate radio 
frequency energy. If it is not installed and used in 
accordance with the instruction manual, it may cause 
interference to other equipment. If this unit causes 
interference (which can be determined by unplugging 
the power cord from the unit), try these measures: 
Relocate the unit with respect to the susceptible 
equipment. Plug this unit and the susceptible 
equipment into different branch circuit.
Consult your dealer. (According to standard EN60601-1-2 
and CISPR11, Class B, Group 1)
WARNING
This symbol indicates the equipotential 
terminal which brings the various parts of a 
system to the same potential.