Agilent Technologies DC122 Benutzerhandbuch

Seite von 55
3.2. 
Channel Input 
The 10-bit digitizer family can be configured with a variety of front-end mezzanine boards offering different 
capabilities. The currently available front end options are detailed below: 
 
 
Front End 
Option 
Impedance Bandwidth 
Guaranteed 
(typical) 
/ Risetime 
 
BW Limiter 
selections 
FS Voltage 
Ranges 
Maximum 
Offset 
Max. Voltage 
Standard 
(Std) 
50 
Ω 
2 GHz 
0.17 ns 
20, 200, 700 MHz 
0.05 – 5.0 V 
± 5 V 
±
 5 V  
50 
Ω 
950 MHz  
( 1 GHz) 
0.35 ns 
20, 200, 700 MHz 
0.05 – 5.0 V  
 
± 5 V 
±
 5 V  
High-
Impedance 
(HZ) 
1 M
Ω 
(300 MHz) 
1.2 ns 
20, 200, 700 MHz 
(see remark below) 
0.05 – 50 V 
 
± 
20 V
 
(> 0.5 V FS) 
± 
200 V
 
(> 5V FS)
  
±
 
300 V
  
High-
Frequency 
(HF) 
50 
Ω 
3 GHz 
0.12 ns 
 1 
± 0.5 V 
±
 
2 V
 
A given unit’s model number will indicate which front end option is installed. If nothing follows the number itself 
then the standard front end option is present. 
Units with either Standard or HF front end options can be ordered with BNC or SMA connectors. The HZ units use 
BNC connectors. 
For FS > 5 V only the 20 MHz Bandwidth Limiter is available. 
3.2.1.  Standard & High Impedance Coupling 
Both AC and DC coupling modes are available. The AC mode couples signals capacitively thus removing the input 
signal’s DC component and filtering out any signal component below 16 Hz for the HZ option or 32 Hz for the Std 
option. DC mode allows all signal components to be passed through to the digitizer. 
3.2.2.  Impedance 
The input channels of the 10-bit Digitizer Family offer termination in either 50 
Ω or 1 MΩ. The 50 Ω coupling mode 
for the Standard and High-Impedance front ends offers high quality termination with better than ± 1% precision. It is 
ideally suited to use with 50 
Ω transmission lines (coax), high bandwidth low impedance (typically 500 Ω) probes or 
active probes. The HF 50 
Ω coupling mode has ± 2% precision The 1 MΩ coupling mode provides a high impedance 
(low load) capability that is suited for use with most standard high impedance probes. The high impedance mode also 
features low (14 pF typical) capacitance that helps to minimize loading effects that can occur when probing high 
frequency circuits. 
3.2.3.  Input Protection 
The input amplifiers are protected against over-voltage signals as shown in the table. The High-impedance limit 
refers to the input signal’s DC + peak AC < 10 KHz voltage. 
3.2.4.  Mezzanine Front-end 
The front-end electronics are all mounted on a removable mezzanine card. In the event of accidental damage or as 
components fatigue over time (e.g. relays in high duty cycle automated testing applications), the mezzanine card 
allows for fast and efficient replacement. 
User Manual: Family of 10-bit Digitizers 
 
Page 28 of 55