Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Static compliance
and resistance
(static C and R)
maneuver
and resistance
(static C and R)
maneuver
Procedure: Patient participation is not required, and patient interference
can affect the accuracy of the maneuver. If the patient initiates a breath
when a maneuver breath is scheduled, the ventilator does not perform the
maneuver. Available in VCV only. Not available in PCV and NPPV.
can affect the accuracy of the maneuver. If the patient initiates a breath
when a maneuver breath is scheduled, the ventilator does not perform the
maneuver. Available in VCV only. Not available in PCV and NPPV.
During the maneuver: The ventilator delivers a square waveform, regardless
of the current setting. No other settings change during the static C and R
maneuver. The maneuver is not performed unless a machine or operator
triggered breath is delivered within 20 seconds. The inspiration time must
be between 200 ms and 5 seconds for the maneuver to occur, and the
maximum plateau duration is 2.5 seconds.
of the current setting. No other settings change during the static C and R
maneuver. The maneuver is not performed unless a machine or operator
triggered breath is delivered within 20 seconds. The inspiration time must
be between 200 ms and 5 seconds for the maneuver to occur, and the
maximum plateau duration is 2.5 seconds.
Results display: Static compliance is displayed in mL/cmH
2
O. Static
resistance is displayed in cmH
2
O/L/s. All compliance and resistance values
are measured on time or operator triggered mandatory breaths only. During
the last saved Static C and R maneuver, Plateau Pressure (P
the last saved Static C and R maneuver, Plateau Pressure (P
plat
) is
measured and displayed as a static value. Static C and R are time-stamped
based on ventilator time and date. Static compliance values from 0 to 350
mL/cmH
based on ventilator time and date. Static compliance values from 0 to 350
mL/cmH
2
O and static resistance values from 0 to 400 cmH
2
O/L/s are
displayed: If either value exceeds its limits or a stable plateau pressure
cannot be achieved, only a message (no data) is displayed.
cannot be achieved, only a message (no data) is displayed.
Range:
Static compliance: 0 to 350 mL/cmH
2
O
Static resistance: 0 to 400 cmH
2
O/L/s
Plateau Pressure: -20 to 130 cmH
2
O
Resolution:
Static compliance: 1 mL/cmH
2
O
Static resistance: 1 cmH
2
O/L/s
Plateau Pressure: 1 cmH
2
O
Accuracy:
Static compliance: ± 1 mL/cmH
2
O ± 20% of true value
Static resistance: ± 3 cmH
2
O/L/s ± 20% of true value
Plateau Pressure: ± (0.5 cmH
2
O) ± (10%) of true value
Inspiratory time
divided by total
breath time (TI/
TTOT).
divided by total
breath time (TI/
TTOT).
Calculated inspiratory time divided by total breath time, or percent
inspiratory time. An eight-breath running average. Available in VCV, PCV,
and NPPV.
inspiratory time. An eight-breath running average. Available in VCV, PCV,
and NPPV.
Applicable to spontaneous and pressure-supported breaths only. If there
were no spontaneous or pressure-supported breaths within the last one
minute, - - - - is displayed.
were no spontaneous or pressure-supported breaths within the last one
minute, - - - - is displayed.
Parameter
Description
Table 13-9: Respiratory Mechanics Maneuver Summary and Data Specifications