4.5 Zero Calibration
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 39
Display
Status...
Main...
Channel
Calib...
TAG
F1
F2
F3
F4
F5
Failures:
No
Maintenance-Requests:
No
Temperature:
20.0 C
Operation:
Ready
0.0
100.0
Range: 1
37.50
ppm CH4
0.00
50.00
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas:
50.00 ppm
Span gas name:
Methane
Start zero calibration procedure !
Expected zero gas:
0.00 ppm
Valves...
Valve position:
Samplegas
Time & Date:
12:50:55 June 16, 1998
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas:
50.00 ppm
Span gas name:
Methane
Start zero calibration procedure !
Expected zero gas:
0.00 ppm
Valves...
Valve position:
Samplegas
Time & Date:
12:50:55 June 16, 1998
⇒
⇒
Change to the
menu "Analyzer
Module Calibration"
menu "Analyzer
Module Calibration"
Press
F5
Caution:
Before starting zero
calibration, realize that
zero gas is available !
(See also section 5.1.1,
p. 5-5... !)
calibration, realize that
zero gas is available !
(See also section 5.1.1,
p. 5-5... !)
Note:
The zeroing of all measure-
ment ranges of the same
channel is running simulta-
neously.
The zeroing of all measure-
ment ranges of the same
channel is running simulta-
neously.
⇒
⇒
Change to the
line "Start zero
calibration
procedure !"
line "Start zero
calibration
procedure !"
Press
once to
get the line "Start zero
calibration procedure !"
white on black.
calibration procedure !"
white on black.
⇒
⇒
Start the zero
calibration
calibration
Press
or