Hanna Instruments hi 504920 Benutzerhandbuch

Seite von 28
3
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your Dealer.
Each meter is supplied complete with:
• HI 62920 Amplified pH electrode with built-in temperature
sensor, DIN connector and 2 m cable
• Hanna HI 504910 interface cable (HI 7920)
• pH 4.01 buffer, 1 sachet (20 mL)
• pH 7.01 buffer, 1 sachet (20 mL)
• Instruction Manual
• 1.5V AA size Alkaline Batteries (4 pcs)
• 12 VDC power adapter
• Rugged Carrying Case.
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in their original packaging together with the sup-
plied accessories.
HI 504920 is a complete portable pH/ORP/temperature meter, to be
used mainly with Hanna HI 504910 Digital Transmitter for two
different applications: it can be a reference pH meter for calibrating
the HI 504910 Digital Transmitter, or it can be set as user interface
for the HI 504910 as well.
All pH measurements are automatically compensated for temperature
(ATC). The instrument housing is made of rugged, lightweight material,
making it truly portable.
Five memorized buffers (4.01, 6.86, 7.01, 9.18 and 10.01 pH) and
wrong buffer recognition technology make calibration simple and error
free. One or two-point calibration can be performed.
An user friendly interface provides clear messages regarding errors,
functions and more.
An alarm time-out is available to alert the user if too much time has
elapsed since the last pH calibration and that re-calibration may be
required.
The meter provides a controlled access to calibration and GLP settings
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION