Setup of the Unit / Setup of E-Mail
30
Setup of the Unit 
Perform the following settings when the unit is used for 
the first time after being purchased, or when the IP 
address assigned to the UP-D77MD is changed.
1
Open the DICOM Settings page.
For detailed information on how to open the page, 
see “Opening the Setup Window” (on page 27).
2
Perform the required settings related to TCP/IP.
1
Enter the IP address of the UP-D77MD on the 
network in the IP Address text box.
2
Enter the sub-net mask in the Sub-net Mask text 
box.
3
Enter the IP address of the default gateway if a 
gateway is used on the network.
3
Enter the DICOM AE title in the AE Title text box.
Up to 16 characters can be entered.
4
Click the [Save settings] button.
The settings are saved.
Setup of E-Mail
The E-Mail settings page allows you to set up E-mail 
service so that an e-mail message will be sent 
automatically when an error occurs in the UP-D77MD.
1
Open the E-Mail settings page.
For detailed information on how to open the page, 
see “Opening the Setup Window” (on page 27).
2
Set criteria determining when an e-mail message 
should be sent.
Place the check marks on the check box 
corresponding to an error that requires notification 
via an e-mail message.
IP Address text box
Sub-net Mask text box
Default Gateway text box
DICOM AE text box
Error item
Content
Motor/Sensor error
When an error occurs in the motor/
sensor system.
Film/Paper 
Jamming
When a paper has jammed.
More than 30,000 
printed with the 
current thermal 
head
When the total number of prints 
with the current thermal head 
exceeds 30,000.
Motor/sensor error check box
Film/Paper Jamming check box
More than 30,000 printed with the 
current thermal head check box