Hanna Instruments hi 98140 Benutzerhandbuch

Seite von 32
9
PARAMETER SETTINGS
• Once the parameter code (and password
if needed) has been entered, the current
value of the selected parameter on the
upper LCD and the parameter code on
the lower LCD will be displayed. The parameter or a part of it will
blink (e.g. only the hour blinks if time has been selected).
• Enter the new value using the arrow keys.
• If there is another part of the parameter to be set (e.g. minutes
for the time), press RANGE to gain access and then enter the new
value using the arrow keys.
• Press ALT and CFM to confirm the value.
• If the entered value is not acceptable, the "WRONG" indication
will be displayed for a few seconds and then the meter will ask for
a correct value.
• If the confirmed value is accepted, the meter will pass to the next
parameter (asking for the password if it is protected).
Note: If ALT and FNC are pressed before parameter value confirma-
tion, the meter will not update the parameter and after
escaping will ask for a new setup code.
The following table lists the setup codes along with the description of
the specific setup items, their valid values and the factory settings
(default):
Code
Valid values
Default
00 Instrument ID code
0000÷9999
0000
01 Current time
1
hh:mm
00:00
02 Current date
1
dd:mm
01:01
03 Current year
1
YYYY
1998
10 Calibration alarm time-out
01÷99 days , OFF
OFF
20 Auto-Off/Power down time-out ON, OFF
ON
21 Firmware version
22 Battery level test
99 Password
2
0000÷9999
0000