PRELIMINARY EXAMINATION
1
Remove the instrument from the packing
material and examine it carefully to make
sure that no damage has occurred during
shipment. If noticeable damage is evident,
notify your Dealer.
HI 955501 comes supplied with:
9V battery
instruction manual.
Note: Save all packing material until you
are sure that the instrument functions
correctly. All defective items must be
returned in their original packing.
GENERAL DESCRIPTION
HI 955501 
HI 955501 
HI 955501 
HI 955501 
HI 955501 allows temperature measurement
using interchangeable PT100 probes.
H
H
H
H
HIIIII 955501
955501
955501
955501
955501 allows temperature measurements
with two different resolutions: 0.1°C between
±199.9 and 1°C outside.
Standard features include low battery detec-
tion and self-explanatory symbols to remind
the user of the operating mode or condition.
Untitled-9
4
FUNCTIONAL DESCRIPTION
AND SPECIFICATIONS
2
Range
-199.9 to 199.9°C
-200 to 850°C
Resolution
0.1°C (±199.9°C)
1°C (outside)
Accuracy
for one year, excluding probe error
Low Range
±0.2°C & ±1 digit (-120 to 200°C)
±1°C & ±1 digit (outside)
High Range
±1°C & ±1 digit (-170 to 450°C)
±1% F.S. & ±1 digit (outside)
Typical EMC
±0.5°C
Deviation
Probe
HI 768 series (not included)
Battery Type
9V
Life
150 hours of continuous use
Environment
 0 to 50°C (32 to 122°F)
95% RH
Dimensions
143x80x38 mm(5.6x3.2x1.5")
Weight
320 g (11.3 oz.)
ins rum n s
RANGE
PT100
THERMOMETER
HI 955501
-199.9°C  to  199.9°C
-200°C  to  850°C
RANGE:
OFF
ON
°C
1
2
3
1. Liquid Crystal Display
2. ON/OFF Key
3. Range Key