eutech-instruments cyanide epoxy Benutzerhandbuch

Seite von 18
Instruction Manual 
 
Cyanide Electrode
 
 
3
 
EUTECH INSTRUMENTS PTE LTD. 
CYANIDE ION ELECTRODE 
INSTRUCTION MANUAL 
 
GENERAL INSTRUCTIONS 
 
Introduction 
 
Eutech Instruments Cyanide Ion Electrode is used to measure cyanide ions in aqueous solutions 
quickly, simply, accurately, and economically. 
 
*IMPORTANT NOTICE* 
 
Acidic cyanide solutions produce hydrogen cyanide (HCN) gas, highly toxic whether breathed or 
absorbed through the skin. Use of the proper and recommended ionic strength adjuster (ISA) will 
keep the solution pH above 10. If measurements in acidic solution are necessary (decomplexing 
procedure as given in the section Complexation),  THIS PROCEDURE MUST BE DONE IN A 
FUME HOOD
.
 
 
Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic. 
 
Required Equipment 
 
1. 
A pH/mV meter or an ion meter, either line operated or portable. 
 
2. 
Semi-logarithmic 4-cycle graph paper for preparing calibration curves when using the 
meter in the mV mode. 
 
3. 
A magnetic stirrer. 
 
4. 
Eutech Cyanide Ion Combination Epoxy-body Electrode, Code no. EC-CN-03. 
 
5. 
Lab-ware made of plastic, not glass. 
 
6. 
Polishing Paper, Code no. EC-MIS-PP, to polish dirty or etched electrode membranes. 
 
Required Solutions 
 
1. 
Deionized or distilled water for solution and standard  preparation. 
 
2. 
Eutech Cyanide Ionic Strength Adjuster (ISA), 10M NaOH, Code no. EC-ISA-CN1-BT. 
To prepare the ISA from your own laboratory stock, fill a 1,000 ml beaker with about 900 
ml of distilled water. While gently stirring the solution under a hood, slowly add 400 
grams of reagent-grade sodium hydroxide. Transfer the solution quantitatively to a one 
liter volumetric flask after the solid NaOH has dissolved and the beaker has cooled. Dilute 
to the mark with distilled water, cap, and upend several times to thoroughly mix the 
solution. Store in a plastic bottle.