2 Safety
 | 
2.3 Safety instructions
2.3 Safety instructions
2.3.1 General
WARNING
Injuries or harm caused by damaged functional components
Damaged functional components can cause personal harm or injury.
▶ If functional components are damaged: Stop work,remedythe problem or inform
the service engineer.
CAUTION
Malfunction due to electromagnetic fields
This product meets all the requirements that are in effect relating to electromagnetic
fields. Due to the complex interference between mobile phones and medical devi-
ces, it is not, however, possible to completely exclude the possibility of the product
being affected by a functioning mobile phone.
▶ Refrain from using mobile phones in the practice and clinic area!
▶ Turn off electronic equipment, such as data storage, hearing aids etc, du-
ringoperation !
CAUTION
Risks caused by electromagnetic fields
Electromagnetic fields may interfere with the functions of implanted systems (such
as pacemakers).
▶ Consult the patient before treatment!
This KaVo product is not permitted for use in areas where there is a risk of explosion.
The following individuals are authorised to repair and maintain the KaVo product:
▪ Engineers from KaVo offices.
▪ Engineers working for KaVo authorised dealers, specially trained by KaVo.
▪ Independent engineers specially trained by KaVo.
2.3.2 Product-specific
CAUTION
Risk of injury due to electric shock
Electric shock
▶ Remove bus covers only when ERGOcom 3 is switched off.
▶ Check the additionally connected cable at earthing connection M . R = < 0.1 Ohm
▶ Observe standards EN 60601-1-1, IEC 950, EN 550022 and EN 550024 in
combination with compatible displays.
▶ In the case of installation in rooms used for medical purposes, this must be
designed in accordance with DIN/VDE 0100-710 (erection of low voltage
systems).
▪ Do not place any additional portable multiple sockets on the floor.
▪ Do not connect any additional portable multiple sockets or extension cables to
the system.
▪ Do not connect any electrical equipment that is not part of the system.