2.2.3
2 - 11
2.2.3 Manual Operation Panel
Figure 2.2-3 shows the manual operation panel.
Figure 2.2-3 shows the manual operation panel.
Fig. 2.2-3 Manual Operation Panel
• Stigma/alignment adjusting section
[Stigma/alignment] [X], [Y] knobs: Use for astigmatism correction in a usual practice manual
mode.
In the electron optics alignment mode, these knobs serve
for alignment adjustment.
In the electron optics alignment mode, these knobs serve
for alignment adjustment.
• Focus adjusting section
[Focus] [Coarse] [Fine] knobs:
Use for coarse and fine focus adjustment.
• Magnification/Image shift control section
[Magnification] knob:
Use for magnification setting.
[Image Shift] knobs:
Move image by electrical image shift.
• Brightness/Contrast adjusting section
[Brightness] knob:
Use for brightness adjustment.
[Contrast] knob:
Use for contrast adjustment.
Brightness/Contrast Adjusting Section
Focus Adjusting Section
Magnification/Image Shift Control Section
Stigma/alignment Adjusting Section