(0070,0001)
Graphic Annotation Sequence
(0070,0005)
>> Graphic Annotation Units
PIXEL
(0070,0020)
>> Graphic Dimensions
2
(0070,0021)
>> Number of Graphic Points
3
(0070,0022)
>> Graphic Data
[values]
(0070,0023)
>> Graphic Type
POLYLINE
(0070,0024)
>> Graphic Filled
-
2.2.1.3.1.3.1.6.3
Rectangle
A rectangle is always provided with a corresponding text (for measurements). If the text is moved by
the operator, an extra text object is added to the GSPS.
the operator, an extra text object is added to the GSPS.
Graphic Layer Module
(0070,0060)
Graphic Layer Sequence
(0070,0002)
>Graphic Layer
ID
(0070,0068)
>Graphic Layer Description
Rectangle or Rectangle_SAL
Graphic Annotation Module
(0070,0001)
Graphic Annotation Sequence
(0070,0002)
> Graphic Layer
[ID of the corresponding layer]
(0070,0008)
> Text Object Sequence
(0070,0003)
>> Bounding box annotation units
PIXEL
(0070,0004)
>> Anchor Point Annotation Units
-
(0070,0006)
>> Unformatted Text Value
[value]
(0070,0010)
>> Bounding Box Top Left Hand Corner
[value]
(0070,0011)
>> Bounding Box Top Right Hand Corner
[value]
(0070,0012)
>> Bounding Box Text Horizontal Justification
LEFT
(0070,0014)
>> Anchor Point
-
(0070,0015)
>> Anchor Point Visibility
-
(0070,0009)
> Graphic Object Sequence
(0070,0005)
>> Graphic Annotation Units
PIXEL
(0070,0020)
>> Graphic Dimensions
2
(0070,0021)
>> Number of Graphic Points
5
(0070,0022)
>> Graphic Data
[values]
(0070,0023)
>> Graphic Type
POLYLINE
(0070,0024)
>> Graphic Filled
N
2.2.1.3.1.3.1.6.4
Circle
A circle is always provided with a corresponding text (for measurements). If the text is moved by the
operator, an extra text object is added to the GSPS.
operator, an extra text object is added to the GSPS.
Graphic Layer Module
(0070,0060)
Graphic Layer Sequence
(0070,0002)
>Graphic Layer
ID