Emerson Process Management CFA-3000 Benutzerhandbuch

Seite von 74
23
CFA3000 MANUAL
SECTION 3
INSTRUMENT DESCRIPTION
2.     The sample display for 20 mA and for 5 Vdc:
0020
2.
To exit, press             .
To display the full­scale concentration embedded in the E­Prom:
Press              two times. To exit, press             .
3.4.1   Record 
mV
Use the 
mV value to determine that the method’s span and slope are correct. The change in millivolts (mV)
is the spread between a baseline and a full­scale standard at the time of calibration. The values are relative to
the chemical parameter being measured
Record the millivolt values for both baseline and full­scale standard.
1.
Press   
The sample Display for value of 150 mV is:     0 1 5 0
2.
To exit function, press             once.
3.4.2   Initiate Auto Calibration
The analyzer is calibrated automatically usually every two days, depending on the chemistry. Auto calibration
performs both baseline and full­scale calibration. Baseline calibration measures a solution with 10% concentra­
tion of the species of interest.  Full­scale calibration measures a standard solution in the upper range of the
parameter of interest.
Before starting calibration, make sure that the analyzer is running reaqgents and baseline
solution for at least 20 minutes.