CFA3000 MANUAL
SECTION 5
TROUBLESHOOTING PROCEDURES
Problem
Probable Cause
Remedy
Alarms not 
activated
Incorrect 
settings
Readjust 
settings, making
sure that low
alarm is set
lower than 
high alarm 
Insufficient time
for alarm to
respond
Alarm condition
must exist for
minimum of 60
seconds before
alarm activates
Signal not 
repetitive
(same sample
not repetitive)
Small leak or
clogging in 
fluidics
Check all tubing
lines and 
connections
Clogged tubing
Replace tubing
No signal, or
signal “off­scale”
high
Connector or
wires loose
Faulty circuit
board
Incorrect mV
value
Check circuit
boards for 
proper seating
Replace power
supply circuit
board
Loss of mV
Sensitivity
LED ISE on
Reagents, 
standards old,
contaminated or
improperly 
prepared
Change
reagents
Reagent straws
in wrong
reagent or
standard
Check 
placement of
straws
Problem
Probable Cause
Remedy
Loss of sample
flow
Replace sample
filter on overflow
panel
Decreased
pumping volume
Tubes pinched
shut or clogged
from deposits
Remove 
condition 
causing
“pinched” lines;
replace sample
filter at overflow
panel
Leak detector
LED lights up
Leak in fluidics
Repair leak and
dry leak detector
contacts; turn
main power
OFF for 
20 seconds
High humidity
Adjust screw on
leak detector
potentiometer
on analog
board; turn
clockwise to
decrease 
sensitivity
Connector faulty
or not properly
insertede
Check 
connection to
analog board;
replace analog
board.
5.2  Test Functions
Table  5­2:  Test Functions
Problem
Probable Cause
Remedy
Front panel 
display
Test LED’s and
LCD on keypad:
a.  Press Power
button and reset
button on the
CPU board (yel­
low ejector)
b.  Release
power, then
Reset button.
Press power 
Press: