Emerson Process Management CFA-3000 Benutzerhandbuch
50
5. If you purchased program from Rosemount
Analytical, run it from your computer, by typing in
CFATERM, then press Return.
6. When a blinking cursor appears on the screen,
press Return. The screen will display M>.
7. At this point, all of the RS232C functions listed in
Table 84 are accessible.
8. To set up a file to collect data in the computer,
press <altL> (formatted) or <altC> (capture).
9. Type in file name (up to 8 characters).
10. To start collecting data, press Return. The display
10. To start collecting data, press Return. The display
will show a blinking cursor.
11. To access the command functions again, press
Return. The screen will display M>.
Use the commands in Table 85 to control the analyzer.
Table 85: RS232C Commands
CFA3000 MANUAL
SECTION 8
ADDIDIONAL FEATURES
Command
Function
AUTOCAL
Runs fullscale
automatic calibration
automatic calibration
AUTOVALV
Resumes automatic valve
selection
selection
DATA
Forces printing of
analyzer data
analyzer data
DISVALV
Command
Displays stream
parameters, including
sample cycles, low and
Function
parameters, including
sample cycles, low and
Function
MANVALV
To operate valves
manually (for multi
stream, one stream
continuously). For
example, to monitor
stream 3, enter
MANVALV 3.
manually (for multi
stream, one stream
continuously). For
example, to monitor
stream 3, enter
MANVALV 3.
RESET
Resets the computer for
a full wipeout
a full wipeout
SETCLO
Sets clock calendar, Date
is entered as follows: day
of week, year, month,
date, hour, minute,
second. Day of week is
coded as
Sunday = 0, Monday = 1,
Tuesday = 2, etc. Year is
entered as last two digits
of year. Month and date
must be two digits, use
leading zeros as
necessary. A 24 hour
format is used for hour
entry. For example, to set
the clock at Thursday,
April 12, 2003 at 4:35
p.m., enter SETCLO
4030412163500.
is entered as follows: day
of week, year, month,
date, hour, minute,
second. Day of week is
coded as
Sunday = 0, Monday = 1,
Tuesday = 2, etc. Year is
entered as last two digits
of year. Month and date
must be two digits, use
leading zeros as
necessary. A 24 hour
format is used for hour
entry. For example, to set
the clock at Thursday,
April 12, 2003 at 4:35
p.m., enter SETCLO
4030412163500.
SETHAL
Changes stream high
alarm setting. For
alarm setting. For
example, to change
stream 1 to 25, enter
SETHAL 1,25.
SETHAL 1,25.
SETLAL
Changes stream low
alarm
setting. For example, to
change stream 2 to 50,
enter SETLAL 2,50.
alarm
setting. For example, to
change stream 2 to 50,
enter SETLAL 2,50.
SETVTIM
Changes stream valve
cycles. For example, to
change stream 2 to 2
cycles, enter
SETVTIM 2,20.
cycles. For example, to
change stream 2 to 2
cycles, enter
SETVTIM 2,20.