5.   If you purchased program from Rosemount 
Analytical, run it from   your computer, by typing in
CFATERM, then press Return.
6. When a blinking cursor appears on the screen,  
press Return.  The screen will display M>.
7.   At this point, all of the RS232­C functions listed in  
Table 8­4 are accessible.
8.   To set up a file to collect data in the computer, 
press <alt­L> (formatted) or <alt­C> (capture).
9.   Type in file name (up to 8 characters).
10. To start collecting data, press Return. The display  
will show a blinking cursor.
11. To access the command functions again, press 
Return. The screen will display M>.
Use the commands in Table 8­5 to control the analyzer.
Table 8­5:  RS232­C Commands
CFA3000 MANUAL
SECTION 8
ADDIDIONAL FEATURES
Command
Function
AUTOCAL
Runs full­scale 
automatic calibration
AUTOVALV 
Resumes automatic valve  
selection
DATA
Forces printing of 
analyzer data
DISVALV 
Command
Displays stream 
parameters, including  
sample cycles, low and  
Function
MANVALV
To operate valves 
manually (for multi­ 
stream, one stream 
continuously). For 
example, to monitor  
stream 3,  enter 
MANVALV 3.
RESET
Resets the computer for   
a full wipeout
SETCLO
Sets clock calendar, Date
is entered as follows: day
of week, year, month,
date, hour, minute,
second. Day of week is
coded as
Sunday = 0, Monday = 1,
Tuesday = 2, etc. Year is
entered as last two digits
of year. Month and date
must be two digits, use
leading zeros as 
necessary. A 24 hour 
format is used for hour
entry. For example, to set
the clock at Thursday,
April 12, 2003 at 4:35
p.m., enter SETCLO
4030412163500.
SETHAL 
Changes stream high
alarm setting. For
example, to change
stream 1 to 25, enter 
SETHAL 1,25.
SETLAL 
Changes stream low
alarm
setting. For example, to
change stream 2 to 50,
enter SETLAL 2,50.
SETVTIM
Changes stream valve
cycles. For example, to
change stream 2 to 2
cycles, enter 
SETVTIM 2,20.