Emerson 2200S User Manual

Page of 158
80
Micro Motion
®
 Model 2200S Transmitters
Testing and Moving to Production
Figure 7-8
ProLink II – mA output trim procedure 
Read mA output at 
receiving device
Read mA output at 
receiving device
Equal?
Enter receiving device 
value in Enter Meas
Read mA output at 
receiving device
Read mA output at 
receiving device
Equal?
Enter receiving device 
value in Enter Meas
4 (12)  mA trim
20 mA trim
Yes
No
Yes
Next
Next
Next
ProLink >
     Calibration >
          Milliamp 1 Trim 
Back
No
Back
Finish
Set mA measurement point