Chapter 9. Diagnostics and Troubleshooting
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
205
E3
Velocity Range - 
Velocity exceeds 
programmed  limits 
(Velocity limit is 
programmed using 
the Service Menu 
Error Limits option 
as described in 
Chapter 8.) 
Programming error, poor 
flow conditions, or bad 
transducer spacing.
Make sure flow rate is within 
±12m/s (±40 ft/s). Check 
programming and transducer 
spacing. If programmed value 
is outside of limit, change the 
limits as described in Chapter 8 
on page 184. See Flowcell Prob-
lems on page 208 and Trans-
ducer Problems on page 210.
E4
Signal Quality
If too high - electronic 
failure. 
If too low - flowcell, 
electrical problem.
Check for source of electrical 
interference. Check electronics 
with a test flowcell. If unit still 
fails to operate, contact the 
factory
E5 
Amplitude Error
Excessive particles or 
bubbles present in fluid. 
A lot of second phase 
present
See Flowcell Problems on 
page 208.
E6
Cycle skip, 
Acceleration
Poor flow conditions or 
bad transducer spacing.
Check transducer spacing as 
programmed and as set on 
pipe. See Flowcell Problems on 
page 208 and Transducer 
Problems on page 210.
E7 
Analog Out Error
Under current output.
Check that output load is within 
specification <550 ohm.
E8
Temperature Input 
Supply
Supply temperature 
input out of range.
Check cable and transmitter.
E9
Temperature Input 
Return
Return temperature 
input out of range.
Check cable and transmitter.
Table 7: Error Messages (cont.)
Error
Problem
Possible Cause
Action