Chapter 3. Programming Site Data
66
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.11.2 Entering Reynolds Correction (cont.)
1.
The first prompt asks if you want to enable the Reynolds Correction 
factor, a number based on the Kinematic Viscosity and flow rate of the 
fluid.
Note:
If you are using clamp-on transducers, you should enable Reynolds 
Correction. 
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press 
[ENTER]
 to confirm your selection.
Note:
If you disable Reynolds Correction, you will not be able to enter any 
other values in this window.
2.
If you enable Reynolds Correction, you will also need to enter the 
Kinematic Viscosity of the fluid (available in the brochure Soundspeeds 
and Pipe Size Data
, 914-004). The prompt asks for a kinematic 
viscosity value.
a.
Press 
[ENTER]
 to open the text box.
b.
Use the numeric keys to enter the desired value (from Soundspeeds 
and Pipe Size Data
).
c.
Press 
[ENTER]
 to confirm your selection.
You have completed entering data for a single Kinematic Viscosity factor. 
Proceed to Entering a Calibration Factor on page 67.
To confirm the entries and return to Operate Mode, press 
[F3]
 (OK). The 
PT878 returns to Operate Mode.
To leave the window without confirming the entries, press 
[F2]
 (Cancel) 
or the 
[ESC]
 key.