Emerson Process Management 3051N User Manual

Page of 100
Reference Manual 
00809-0100-4808, Rev CA
June 2008
5-11
Rosemount 3051N
NOTE
Mounting Bracket option code must be specified last in the Rosemount model 
number even if optional meter is ordered.
CONFIGURATION 
INFORMATION
Transmitter is shipped as follows (unless different calibration is specified).
Engineering Units
Code
Housing Material
Conduit Entry Size
A
Polyurethane-covered Aluminum
½–14 NPT
J
SST
½–14 NPT
Code
Alternate Flange Options (Requires Materials of Construction Code 0)
H2
Traditional Flange, 316 SST, SST Drain/Vent, SST Flange Adapter
Code
Mounting Bracket Options
B2
Traditional Flange Bracket for Panel Mounting, CS Bolts
B4
Coplanar Flange Bracket for Panel Mounting, all SST
BS
Universal Traditional Flange Bracket for Panel Mounting (SST), CS Bolts
PM
Traditional Flange Bracket for Pipe Mounting, all SST
Code
Meters (optional)
M5
LCD Meter for Aluminum Housing (Housing Code A only)
M6
LCD Meter for SST Housing (Housing Code J only)
Typical Rosemount Number: 
3051ND
2
A
2
2
A
1
A
B4
(1) All Rosemount 3051N transmitters are provided as standard with transient protection block (TI) and cleaning for < 1 PPM chloride. 
(2) Rosemount 3051ND0 is available only with Process Flange Code 0 (Alternate Flange H2), O-ring Code A, and stainless steel process flange bolting.
(3) EPR process o-ring is limited to 2000 psi maximum working pressure and 150 °F (66 °C) process temperature.
Differential/Gage:
in H
2
0 (Range 0, 1, 2, and 3) 
psi (Range 4 and 5)
Absolute:
psi (all ranges)
4 mA:
0 (engineering units above)
20 mA:
Upper range limit
Output:
Linear
Flange type:
Specified Rosemount code option
Flange material:
Specified Rosemount code option
O-ring material:
Specified Rosemount code option
Drain/vent:
Specified Rosemount code option
Integral meter:
Installed or none
Alarm:
Upscale
Software tag:
(blank)