Reference Manual
00809-0100-4808, Rev CA
June 2008
A-15
Rosemount 3051N
Add item for ALL 
device types or only 
for this ONE device 
type.
Asks the user whether the hot key item being added should be 
added for all device types or only for the type of device that is 
connected.
Command Not 
Implemented
The connected device does not support this function.
Communication 
Error
The communicator and the device are not communicating 
correctly. Check all connections between the communicator 
and the device and resend the information.
Configuration 
memory not 
compatible with 
connected device
The configuration stored in memory is incompatible with the 
device to which a transfer has been requested.
CPU board not 
initialized ON
The electronics board is not initialized. Return to Rosemount 
Nuclear Instruments, Inc. for repair.
CPU EEPROM write 
failure ON
Message sent to electronics board from HART signal failed. 
Return to Rosemount Nuclear Instruments, Inc. for repair.
Device Busy
The connected device is busy performing another task.
Device 
Disconnected
The device failed to respond to a command. Check all 
connections between the communicator and the device and 
resend the command.
Device write 
protected
Device is in write-protect mode. Data can not be written.
Device write 
protected. Do you 
still want to 
shut off?
Device is in write-protect mode. Press YES to turn the HART 
communicator off and lose the unsent data.
Display value of 
variable on hotkey 
menu?
Asks whether the value of the variable should be displayed 
adjacent to its label on the hotkey menu if the item being added 
to the hotkey menu is a variable.
Download data from 
configuration 
memory to device
Press the SEND softkey to transfer information from the 
communicator memory to the device.
Exceed field width
Indicates that the field width for the current arithmetic variable 
exceeds the device-specified description edit format.
Exceed precision
Indicates that the precision for the current arithmetic variable 
exceeds the device-specified description edit format.
Ignore next 50 
occurrences of 
status?
Select YES to ignore the next 50 occurrences of device status, 
or select no to display every occurrence.
Illegal character
An invalid character for the variable type was entered.
Illegal date
The day portion of the date is invalid.
Illegal month
The month portion of the date is invalid.
Illegal year
The year portion of the date is invalid.
Incompatible CPU 
board and module 
ON
Return to Rosemount Nuclear Instruments, Inc. for repair.
Incomplete 
exponent
The exponent of a scientific notation floating point variable is 
incomplete.
Incomplete field
The value entered is not complete for the variable type.
Message