Agilent Technologies E6380-90073 User Manual

Page of 352
Chapter 1
215
GPIB Command Dictionary
MEASure subsystem
:AFRequency:FM <meas cmnd>
:AFRequency:FM?
These commands set/query the FM deviation measurement. This 
measurement is active when FM Mod or FM Demod is selected by using 
the AFAN:INP command (see page  59) and SNR is not selected by the 
MEAS:AFR:SEL command (see page  216).
Syntax
MEAS:AFR:FM <meas cmd> 
MEAS:AFR:FM? !returns real value
Example
MEAS:AFR:FM:AVERage:RESet !resets the number of averages
MEAS:AFR:FM? 
!returns the measured value
Screen/field equivalent
MEAS:AFR:FM? reads the 
FM Deviation
 measurement field on the AF 
ANALYZER screen when the 
AF Anl In
 field is set to 
FM Mod
, or 
FM 
Demod
. This measurement is also displayed on the RF GENERATOR 
and RF ANALYZER screens.
:AFRequency:FREQuency <meas cmnd>
:AFRequency:FREQuency?
These commands set/query the audio frequency measurement. This 
measurement is active when AF Freq is selected by using the 
MEAS:AFR:SEL ‘AF Freq’ command (see page  216).
Syntax
MEAS:AFR:FREQ <meas cmd> 
MEAS:AFR:FREQ? !returns real value
Example
MEAS:AFR:SEL ‘AF Freq’ !selects the measurement 
MEAS:AFR:FREQ? 
!returns the measurement’s value
Screen/field equivalent
MEAS:AFR:FREQ? reads the 
AF Freq
 measurement field on the AF 
ANALYZER screen. This measurement is also displayed on the RF 
GENERATOR and RF ANALYZER screens.