Agilent Technologies E6380-90073 User Manual

Page of 352
266
Chapter 1
 
GPIB Command Dictionary
MEASure subsystem
:DEC:NAMPs|NTACs:RECC:DATA?
This command reads the decoded RECC data, serially, as it received.
Syntax
MEAS:DEC:NAMP:RECC:DATA? !returns a quoted string
Screen/field equivalent
MEAS:DEC:NAMP:CDAT? reads the 
RECC Data (hex)
 measurement 
field on the SIGNALING DECODER screen when the 
Mode
 field is set 
to 
NAMP-NTAC
 , the 
Measure 
field is set to
 Data 
and the 
Channel 
field is 
set to
 Cntl
.
:DEC:NAMPs|NTACs:RVC:DATA?
This command reads the decoded RVC data, serially, as it is received. 
Syntax
MEAS:DEC:NAMP:RVC:DATA? !returns a quoted string
Screen/field equivalent
MEAS:DEC:NAMP:RVC:DATA? reads the 
RVC Data (hex)
 
measurement field on the SIGNALING DECODER screen when the 
Mode
 field is set to 
NAMP-NTAC
, the 
Measure
 field is set to 
Data
 and the 
Channel
 field is set to 
Voice
.
:DEC:NAMPs|NTACs:DSAT:DATA?
This command reads the decoded DSAT/DST data, serially, as it is 
received.
Syntax
MEAS:DEC:NAMP:DSAT:DATA? !returns a quoted string
Screen/field equivalent
MEAS:DEC:NAMP:DSAT:DATA? reads the 
DSAT/DST (hex)
   
measurement field on the SIGNALING DECODER screen when the 
Mode
 field is set to 
NAMP-NTAC
, the 
Measure 
field is set to
 DSAT 
and the 
Channel 
field is set to
 Voice
.