Princeton 4411-0106 User Manual

Page of 136
10 
PIXIS System Manual 
Version 2.C 
 
Grounding and Safety 
Before turning on the power supply (air-cooled system or liquid-cooled system with a 
CoolCUBE
II
 circulator), the ground prong of the power cord plug must be properly 
connected to the ground connector of the wall outlet. The wall outlet must have a third 
prong, or must be properly connected to an adapter that complies with these safety 
requirements.  
If the equipment is damaged, the protective grounding could be disconnected. Do not use 
damaged equipment until its safety has been verified by authorized personnel. 
Disconnecting the protective earth terminal, inside or outside the apparatus, or any 
tampering with its operation is also prohibited. 
Inspect the supplied power cord. If it is not compatible with the power socket, replace the 
cord with one that has suitable connectors on both ends.  
Replacement power cords or power plugs must have the same polarity and power rating 
as that of the original ones to avoid hazard due to electrical shock.  
Precautions 
To prevent permanently damaging the system, please observe the following precautions: 
  The CCD array is very sensitive to static electricity. Touching the CCD can 
destroy it. Operations requiring contact with the device can only be performed at 
the factory.  
  If you are using high-voltage equipment (such as an arc lamp) with your camera 
system, be sure to turn the camera power ON LAST and turn the camera power 
OFF FIRST. 
  Use caution when triggering high-current switching devices (such as an arc lamp) 
near your system. The CCD can be permanently damaged by transient voltage 
spikes. If electrically noisy devices are present, an isolated, conditioned power 
line or dedicated isolation transformer is highly recommended. 
  Do not block air vents on the camera. Preventing the free flow of air overheats 
the camera and may damage it. 
  If the PIXIS camera system is used in a manner not specified by Princeton 
Instruments, the protection provided by the equipment may be impaired. 
UV Coating 
If you have a camera with a UV (Lumogen or Unichrome) coated CCD, protect it from 
unnecessary exposure to UV radiation. This radiation slowly bleaches the coating, 
reducing sensitivity.  
WARNING!
 
WARNING!
 
Caution