Princeton 4411-0106 User Manual

Page of 136
Appendix D 
Spectrograph Adapters 
115 
 
115 
Acton Series Spectrograph (PIXIS with 3.60/3.88 Bolt Circles)  
 
 
 
Qty 
P/N 
Description 
1. 
2826-0127  
Screw, 10-32 
 1/4, Button Head Allen Hex, Stainless Steel 
Assembly Instructions 
1.  Make sure that the shipping cover has been removed from the detector port on the 
spectrograph. 
2.  Loosen the setscrews holding the sliding tube in the spectrograph. 
3.  Rotate the sliding tube as you remove it from the spectrograph.  
4. 
Remove the spacer plate from the sliding tube by removing the three (3) socket head 
screws.
 
5.  Mount the sliding tube to the detector nose with the three (3) 1/4" long button head 
screws. 
6.  Rotate the sliding tube as you gently insert it into the spectrograph. 
7.  Secure the sliding tube with the setscrews. 
Note:  Adapter parts are machined to provide a tight fit. If you need to remove the sliding 
tube from the spectrograph, first loosen the two setscrews that secure it, and then rotate the 
tube as you pull it out. If you have removed the sliding tube from the spectrograph, rotate the 
sliding tube as you re-insert it, and tighten the setscrews afterwards to secure it. Forcing the 
tube into the spectrograph could permanently damage the tube and the spectrograph opening.