Princeton 4411-0106 User Manual

Page of 136
Appendix D 
Spectrograph Adapters 
119 
 
119 
IsoPlane SCT-320 (PIXIS with Flange) 
 
 
 
Qty 
P/N 
Description 
1. 
2826-0120 
Screw, 10-32 
 1/2, Hex Head, Stainless Steel 
Assembly Instructions 
1.  Make sure that the shipping cover has been removed from the detector mounting 
plate on the IsoPlane. 
2.  Leaving 1/4" of thread exposed, screw the three (3) hex head screws into the 
mounting plate. 
3.  Mount the detector to the mounting plate: the text should be right-reading on the back 
of the detector.  
4.  Tighten the three screws with a 5/16” open end wrench. 
Notes:  
1.  Rotational alignment of the detector to the spectrograph optics is done by loosening 
and subsequently retightening the screws at the mounting plate corners. The holes are 
slotted to allow about 4° of rotation. 
2.  If the IsoPlane was ordered with an internal shutter at the entrance slit, the PIXIS, 
when connected to the Shutter input on the IsoPlane, can control that shutter. If the 
PIXIS is using its own internal shutter, the IsoPlane shutter will need to be removed. 
See the IsoPlane SCT 320 manual for shutter removal instructions.