Princeton 4411-0106 User Manual

Page of 136
 
 
19 
 
Chapter 3 
Initial System Verification 
The  list and diagrams  below  briefly  describe the  sequence  of  actions required to 
install your system and prepare to gather data. Refer to the indicated references for 
more detailed information. 
Action 
Reference 
1.  If the system components have not already been unpacked, unpack 
them and inspect their carton(s) and the system components for 
in-transit damage.  
Chapter 4 System Setup, 
page 23 
2.  Verify that all system components have been received. 
Chapter 4 System Setup, 
page 24 
3.  If the components show no signs of damage, verify that the 
appropriate power cord has been supplied with the power supply. 
Chapter 4 System Setup, 
page 24 
4.  If the application software is not already installed in the host 
computer, install it.  
Chapter 4 System Setup, 
page 26 & Software manual 
5.  If a USB 2 interface card is not already installed in the host 
computer, install it. Follow the manufacturer’s instructions. 
 
6.  Depending on application, attach lens to the camera or mount the 
camera to a spectrograph. 
Chapter 4 System Setup, 
page 29 
7.  With the power supply disconnected from the camera, connect the USB 
cable to the USB port at the rear of the camera and to the USB port at 
the computer. 
 
8.  Air-Cooled System: Plug the power supply into the rear of the 
camera and plug the power supply into the power source.  
Liquid-Cooled System: Plug the power supply into the rear of the 
camera and plug the power supply into the power source. Make the 
hose connections to the camera. Plug the circulator into the power 
source. Add coolant if necessary. Turn on the circulator. 
 
 
Chapter 4 System Setup, 
page 27 
9.  Turn the camera ON. 
 
10.  Turn on the computer and begin running the application software. 
Software manual 
11.  Enter the hardware setup information or load the defaults from the 
camera. 
Chapter 5 Operation,  
page 29 
12.  Set the target array temperature. 
Chapter 5 Operation, 
page 57 
13.  When the system reaches temperature lock, wait an additional 20 
minutes and then begin acquiring data in focus mode.  
Chapter 5 Operation, 
page 38 or page 41