  Check the brightest regions of the image to determine if the A/D converter is at 
full-scale. A 16-bit A/D is at full scale when the brightest parts of the image 
reach an intensity of 65535. 
Adjust the aperture to where it is just slightly 
smaller (higher f-stop) than the setting where maximum brightness on any part 
of the image occurs.  
  Set the focus adjustment of the lens for maximum sharpness in the viewed 
image. 
5.  After you have focused the camera, you can stop Preview mode, continue 
Preview mode, or begin Acquire mode. 
Spectroscopy 
The following paragraphs provide step-by-step instructions for operating a PIXIS in a 
spectroscopy setup in LightField for the first time. The intent of this simple procedure is 
to help you gain basic familiarity with the operation of your system and to show that it is 
functioning properly. Once basic familiarity has been established, then operation with 
other operating configurations, ones with more complex timing modes, can be performed. 
An underlying assumption for the procedure is that the camera is to be operated with a 
spectrograph (such as an Acton Series 2300 spectrograph) on which it has been properly 
installed (refer to Appendix D for instructions for mounting a spectrograph adapter to the 
PIXIS). A suitable light source, such as a mercury pen-ray lamp, should be mounted in 
front of the entrance slit of the spectrograph. Any light source with line output can be 
used. Standard fluorescent overhead lamps have good calibration lines as well. If there 
are no "line" sources available, it is possible to use a broadband source such as tungsten 
for the alignment. If this is the case, use a wavelength setting of 0.0 nm for alignment 
purposes. 
Overexposure Protection: Cameras that are exposed to room light or other 
continuous light sources will quickly become saturated. If the camera is mounted to a 
spectrograph, close the entrance slit of the spectrograph to reduce the incident light.  
Assumptions 
The following procedure assumes that  
1.  You have already set up your system in accordance with the instructions in 
Chapter 4. 
2.  You have read the previous sections of this chapter. 
3.  You are familiar with the application software. 
4.  The system is being operated in spectroscopy mode. 
5.  The PIXIS does not have an internal shutter. 
Caution