Chapter 5 
Operation 
51 
 
51 
Confirming the Setup 
1.  Turn on the light source at the spectrograph entrance slit. 
2.  Click on 
Run
 to begin previewing the data. Depending on the display settings, 
you should see either a spectral band (image) or a graph. Background noise will 
decrease as the camera cools to its default temperature. 
3.  Turn off the light source. The data display should change to a background noise 
pattern or low intensity graph. If this occurs you have confirmed that light 
entering the spectrograph is being seen by the camera. Skip Steps 4-5 and 
continue to the “Focusing” topic on page 52. 
4.  If there is little or no difference between the data displayed when the light source 
is on or off: 
a.  Verify that the light source has power and is turned on. 
b.  Verify that the entrance slit is open at least 10 µm.  
c.  Check the 
Exposure Time
 (
Common Acquisition Settings
 expander
)
.  
d.  Confirm that 
Shutter Mode
 is set to Normal (
Shutter
 expander
)
.  
e.  Check the shutter cable connections. 
f.  Verify shutter operation. You should hear the shutter open and close while 
Run
 is active.  
  If you hear a shutter operating and you have done Steps a-e, turn the light 
source on, wait a minute and then turn the light off while you view the 
data display. If the problem is fixed, stop acquisition or continue to the 
Focusing” topic on page 42 ; otherwise, stop data acquisition and 
continue to Step g.  
  If you do not hear a shutter operating and you have done Steps a-e, stop 
data acquisition and continue to Step g. 
g.  Make sure the spectrograph has an entrance slit shutter. An externally 
mounted shutter is easily confirmed. Verifying an internally mounted shutter 
requires access to the inside of the spectrograph: refer to the spectrograph 
manual for instructions. 
h.  Check to see if the PIXIS has a shutter. Loosen and remove the camera from the 
spectrograph. If you see the CCD when you look into the front of the camera, the 
camera does not have an internal shutter. Re-mount the camera to the 
spectrograph. 
i.  If the camera and spectrograph both have shutters, permanent damage to the 
shutter driver may occur. You can either disconnect and remove the entrance 
slit shutter or control it via a Princeton Instruments SHC-EXT Control Box. 
To disconnect and remove an entrance slit shutter: 
  External Slit Shutter: Disconnect the shutter cable and remove the 
shutter assembly from the spectrograph entrance port. 
  Internal Slit Shutter: Disconnect the shutter cable. Removing an 
internally mounted shutter requires access to the inside of the 
spectrograph: refer to the spectrograph manual for removal instructions. 
5.  If you need more help, contact Customer Support. See page 132 for contact 