Princeton 4411-0106 User Manual

Page of 136
Chapter 5 
Operation 
55 
 
55 
IsoPlane SCT-320 Spectrograph 
Because the PIXIS is mounted directly to the mounting plate on the IsoPlane, focusing 
and alignment is different from the way that focusing and alignment are performed for an 
Acton Series spectrograph. The following information assumes that you are familiar with 
the locations of the mounting plate, Micrometer Compartment, and the locking set screw. 
If not, refer to the IsoPlane manual supplied with the spectrograph. 
Note: If the PIXIS has an internal shutter, DO NOT USE the Shutter connector to drive 
a shutter internal to an IsoPlane. If both the camera and spectrograph have shutters, the 
IsoPlane shutter will have to be removed or a Princeton Instruments SHC-EXT Shutter 
Control Box can be used to control the opening and closing of the IsoPlane’s internal 
shutter. See the IsoPlane SCT 320 manual for more information. 
1.  Mount an Acton light source such as the dual HG/NeAr source in front of the 
entrance slit of the spectrograph.  
2.  With the IsoPlane properly connected to the computer, turn the power on, wait for the 
spectrograph to initialize.  
3.  With the PIXIS mounted to the spectrograph and connected to the computer, turn on 
the power and wait for the detector to initialize. 
4.  Start the application software. Because you are using LightField, you will need to drag 
the icons for the PIXIS and the IsoPlane into the 
Experiment Devices
 area.  
5.  Set the spectrograph to 500 nm if using a mercury source or to 0.0 nm if using a 
broadband source.  
6.  Remove the cover from the Micrometer Compartment. 
7.  Using a 3/32” hex wrench, loosen the locking set screw. 
8.  Wait until the detector temperature locks at its default temperature. 
9.  Click on 
Run
 and while continuously acquiring data, adjust the micrometer until you 
maximize the intensity level of a selected peak or peaks. 
10.  Tighten down the locking set screw. 
11.  Place the Micrometer Cover on the spectrograph. Replace and tighten all of the cover 
screws. 
12.  Next adjust the rotation. First, use a 9/64” hex wrench to loosen the four screws at the 
corners of the detector mounting plate. While watching a live display of the 
spectrum, choose a peak to monitor and then rotate the detector (up to 4 degrees of 
rotation are possible). The peak will go from broad to narrow and back to broad. 
Leave the detector rotation set for the narrowest achievable peak.  
Note: With LightField, you can use the 
Align Spectrometer
 
function. 
Alternatively, you can acquire an image, display the horizontal and vertical cursor 
bars, and compare the vertical bar to the line shape on the screen. Rotate the detector 
until the line shape on the screen is parallel with the vertical bar. 
13.  After completing the rotational alignment, re-tighten the four mounting plate screws 
and stop acquisition.