Princeton 4411-0106 User Manual

Page of 136
92 
PIXIS System Manual 
Version 2.C 
 
Serial violations have occurred. Check interface cable. 
 
Figure 57.  Serial Violations Have Occurred dialog 
This error message dialog will appear if you try to acquire an image or focus the camera 
and either (or both) of the following conditions exists:  
  The camera system is not turned ON.  
  There is no communication between the camera and the host computer. 
To correct the problem: 
1.  Turn OFF the camera system (if it is not already OFF).  
2.  Make sure the computer interface cable is secured at both ends. 
3.  After making sure that the cable is connected, turn the camera system power ON
4.  Click 
OK
 on the error message dialog and retry acquiring an image or running in 
focus mode.  
Note: This error message will also be displayed if you turn the camera system OFF or a 
cable comes loose while the WinX application software is running in Focus mode. 
Shutter Failure 
See "Overexposed or Smeared Images", page 90.