Teledyne MODEL T703 User Manual

Page of 287
Teledyne API T703 Calibrator Operation Manual 
Theory of Operation 
 
171 
5 – 6 Seconds 
Analyzer measures the average UV light intensity of Non-O
3
 bearing Sample Gas
 (I
0
)
 
during this period
.
 
CYCLE REPEAT EVERY 6 SECONDS 
PHOTOMETER
 PRESSURE SENSOR
O
3
GEN / PHOTOMETER 
PRESSURE / FLOW SENSOR PCA
O
3
GAS INPUT 
PRESSURE SENSOR
O
3
FLO
SENSO
R
O
3
Gen
era
to
r A
ss
em
bly
O
3
G
E
NERAT
OR
Fl
ow
 C
ontrol 
(1
.0
 LPM)
REF/MEAS
Valve
On Back Panel
Chassis
GAS OUTPUT MANIFOLD
PHOTOMETER 
OUTLET
TO ANALYZER
VENT
TO ANALYZER
PHOTOMETER 
INLET
EXHAUST
PHOTOMETER 
ZERO OUT
PHOTOMETER 
ZERO IN
PHOTOMETER BENCH
PUMP
INTERNAL
VENT
Fl
ow
 C
ontrol 
(5
.0
 lpm)
Pressure 
Regulator
Filter
DRY AIR 
IN
ZERO AIR 
IN
PUMP
CHA
R
C
O
A
L
 
S
C
R
U
BBE
R
blk
blk
blu
blu
orn
orn
orn
orn
red
red
pur
pur
grn
grn
yel
yel
Flow C
ontrol 
(100
 c
m
3
/mi
n
)
Flow Control 
(800 cm
3
/min)
 
Figure 9-17: 
O
3
 Photometer Gas Flow – Measure Cycle 
 
07223B DCN6378