Teledyne GFC 7001E User Manual

Page of 321
Optional Hardware and Software  
 
Model GFC7001E Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
81 
 
CO
2
 gas.  The software is flexible enough to allow you to configure the alarms so that you can have 2 alarm 
levels for each gas.   
 
CO Alarm 1 = 20 PPM 
CO Alarm 2 = 100 PPM 
CO
2
 Alarm 1 = 20 PPM 
CO
2
 Alarm 2 = 100 PPM 
 
In this example, CO Alarm 1 & CO
2
 Alarm 1 will both be associated with the “Alarm 2” relay on the rear panel.  
This allows you do have multiple alarm levels for individual gasses.   
A more likely configuration for this would be to put one gas on the “Alarm 1” relay & the other gas on the “Alarm 
2” relay. 
 
CO Alarm 1 = 20 PPM 
CO Alarm 2 = Disabled 
CO
2
 Alarm 1 = Disabled 
CO
2
 Alarm 2 = 100 PPM 
“Alarm 4” Relay 
 
This relay is connected to the “range bit”.  If the instrument is configured for “Auto Range” & the 
instrument goes up into the high range, it will turn this relay on.