Advanced Features 
 
125 
 
7.2. SETUP  MORE  VARS: INTERNAL VARIABLES (VARS) 
The M300E/EM has several user-adjustable software variables, which define certain operational parameters.  
Usually, these variables are automatically set by the instrument’s firmware, but can be manually redefined using 
the VARS menu.   
The following table lists all variables that are available within the 818 password protected level.  See Appendix 
A-2 for a detailed listing of all of the M300E/EM variables that are accessible through the remote interface. 
Table 7-4:  Variable Names (VARS) 
NO. VARIABLE 
DESCRIPTION 
ALLOWED 
VALUES 
VARS 
DEFAULT 
VALUES 
DAS_HOLD_OFF 
Changes the Internal Data Acquisition System (iDAS) 
HOLDOFF timer. 
No data is stored in the iDAS channels during situations 
when the software considers the data to be questionable 
such as during warm-up or just after the instrument returns 
from one of its calibration modes to SAMPLE Mode.   
May be set for 
intervals 
between  
0.5 – 20 min 
 
15 min. 
CONC_PRECISION 
Allows the user to set the number of significant digits to the 
right of the decimal point display of concentration and stability 
values. 
AUTO, 1, 2,  
3, 4 
AUTO 
DYN_ZERO
 1  
Dynamic zero automatically adjusts offset and slope of the 
CO response when performing a zero point calibration during 
an AutoCal (see Section 9.4). 
ON/OFF 
OFF 
DYN_SPAN 
1  
Dynamic span automatically adjusts the offsets and slopes of 
the CO response when performing a slope calibration during 
an AutoCal  (see Section 9.4)
ON/OFF 
OFF 
CLOCK_ADJ
 
Adjusts the speed of the analyzer’s clock.  Choose the + 
sign if the clock is too slow, choose the - sign if the clock is 
too fast. 
-60 to +60 
s/day 
0 sec 
STABIL_GAS
Selects which gas measurement is displayed when the STABIL
test function is selected. 
CO; CO
2
 & O
2
 
CO 
1
  Use of the DYN_ZERO and DYN_SPAN features are not allowed for applications requiring EPA equivalency. 
2
 
 This VARS only appears if either the optional O
2
 or CO
2
 sensors are installed. 
 
 