Teledyne M300EM User Manual

Page of 398
Specifications and Approvals 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
28 
 
2.2. EPA EQUIVALENCY DESIGNATION 
Teledyne API’s M300E Carbon Monoxide Analyzer is designated as Reference Method Number EQOA-0992-
087 as defined in 40 CFR Part 53, when operated under the following conditions: 
  Range: Any range from 10 ppm to 50 ppm. 
  Ambient temperature range of 10 to 40C. 
  Line voltage range of 90 – 127 and 200 – 230 VAC, 50/60 Hz. 
  Sample filter: Equipped with PTFE filter element in the internal filter assembly. 
  Sample flow of 800  80 cm
3
/min at sea level. 
  Internal sample pump. 
 Software settings: 
Dilution factor 
1.0 
AutoCal 
ON or OFF 
Dynamic Zero 
ON or OFF 
Dynamic Span 
OFF 
Dual range 
ON or OFF 
Auto range 
ON or OFF 
Temp/Pres compensation 
ON 
 
Under the designation, the analyzer may be operated with or without the following options: 
  Rack mount with slides. 
  Rack mount without slides, ears only. 
  Zero/span valve options. 
  Option 50A – Sample/Cal valves, or; 
  Option 50B – Sample/Cal valves with span shutoff & flow control. 
  Internal zero/span (IZS) option with either: 
  Option 51A – Sample/Cal valves, or;   
  Option 51C – Sample/Cal valves with span shutoff & flow control. 
 Status outputs. 
 Control inputs. 
 RS-232 output. 
 Ethernet output. 
  4-20mA, isolated output. 
04288D DCN5752