8.5.5. LCD Display Module  
Verify the functioning of the front panel display by observing it when power is 
applied to the instrument. Assuming that there are no wiring problems and that 
the DC power supplies are operating properly, the display screen should light and 
show the splash screen with logo and other indications of its state as the CPU 
goes through its initialization process. 
8.5.6. Relay Board 
The relay board PCA (04135) can be most easily checked by observing the 
condition of the its status LEDs on the relay board, as described in Section 
8.1.4.3, and the associated output when toggled on and off through signal I/O 
function in the diagnostic menu, see Section 8.1.3.   
1.  If the front panel display responds to button presses and D1 on the relay 
board is not flashing, then either the I
2
C connection between the 
motherboard and the relay board is bad, or the relay board itself is bad. 
2.  If D1 on the relay board is flashing and the status indicator for the output in 
question (heater power, valve drive, etc.) toggles properly using the signal 
I/O function, then the associated control device on the relay board is bad.  
Several of the control devices are in sockets and can be easily replaced.  
The table below lists the control device associated with a particular function: 
Table 8-8:   Relay Board Control Devices 
FUNCTION 
CONTROL 
DEVICE 
IN SOCKET 
Wheel Heater 
K1 
Yes 
Bench Heater 
K2 
Yes 
Spare AC Control 
K3 
Yes 
IZS Valves 
U4 
Yes 
IR Source Drive 
U5 
No 
The IR source drive output can be verified by measuring the voltage at J16 with 
the IR source disconnected.  It should be 11.5± 0.5 VDC. 
8.5.7. Sensor Assembly 
8.5.7.1. SYNC/DEMODULATOR ASSEMBLY 
To verify that the Sync/Demodulator Assembly is working follow the procedure 
below: 
1.  Verify that D1 and D2 are flashing (they flash at different rates, see Table 
  If not check the opto pickup assembly, Section 8.5.7.2 and the GF
  If the wheel drive and opto pickup are working properly then verify that 
there is 2.4 ±0.1 VAC and 2.5 ±0.15 VDC between digital ground and TP 
5 on the sync demod board.  If not then check the wiring between the 
07272B DCN6552