Teledyne API Model T360/T360M Operation Manual 
Getting Started 
 
35 
3.5. ELECTRICAL CONNECTIONS 
 
WARNING 
Never disconnect PCAs, wiring harnesses or electronic subassemblies 
while the instrument is under power. 
NOTE  
To maintain compliance with EMC standards, it is required that the cable length be no greater than 3 
meters for all I/O connections, which include Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, 
USB, RS-232, and RS-485.   
Refer to Figure 3-4 for the locations of the rear panel connections. 
3.5.1. Power Connection 
 
 
The T360 analyzer can be configured for both 100-130 V and 210-240 V at either 
50 or 60 Hz.  To avoid damage to your analyzer, make sure that the AC power 
voltage matches the voltage indicated on the rear panel serial number label and 
that the frequency is between 47 and 63 Hz.   
Attach the power cord to the analyzer and plug it into a power outlet capable of 
carrying at least 10 A current at your AC voltage and that it is equipped with a 
functioning earth ground. 
 
WARNING 
Power connection must have functioning ground connection. 
Do not defeat the ground wire on power plug. 
Turn off analyzer power before disconnecting or  
connecting electrical subassemblies. 
Do not operate with cover off. 
 
CAUTION 
Check the voltage and frequency specifications on the rear panel Model 
Label for compatibility with the local power before plugging the analyzer 
into line power. 
Do not plug in the power cord if the voltage or frequency is incorrect. 