Compare Excel Meter Configurations (Logs/Reports Menu)
Section 5: Startup and meter configuration
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
Ultrasonic Meter Archive log types
There are up to five different types of logs stored in the meter depending on the type of meter 
which are described below.
5.10.2
Compare Excel Meter Configurations (Logs/Reports Menu)
This utility provides the ability to compare configurations stored in Excel® Maintenance logs, 
Archive Logs, Flow Calibration Logs, and Zero Flow Calibration logs. The default location for 
saving the compare results will be the directory from which the first file selected to compare is 
located. This utility can not compare configurations stored in different system units. (i.e. U.S 
Customary vs. Metric). Only Excel® log files can be compared.
Daniel MeterLink will prompt you to select the files to compare. Once the files are selected, a 
dialog will display providing options to further customize the compare operation.
Table 5-26  Ultrasonic Meter Archive Log Types
Log type
Description
Daily Logs 
The ultrasonic meter stores a log record in the Daily Log memory once per day at the 
Contract hour for daily log set under Field Setup Wizard. The Daily Log has a readable semi-
fixed configuration that stores the volumes and associated data accumulated over the past 
24 hours.
For Daniel Mark III Gas Ultrasonic meters, the data registers stored in the Daily logs are listed 
in 
. The ultrasonic meter can store 365 Daily Log records. 
Hourly Logs 
The ultrasonic meter stores a log record in the Hourly Log memory every hour at the top of 
the hour. The Hourly Log has a readable semi-fixed configuration that stores the volumes 
and associated data accumulated over the past hour. 
For Daniel Mark III Gas Ultrasonic meters, the data registers stored in the Hourly logs are 
listed in 
. The ultrasonic meter can store 2400 Hourly Log records. 
 Audit Logs 
For Daniel Mark III Gas Ultrasonic meters, the parameters which affect the gas flow mea-
surement are listed in 
. The ultrasonic meter can store 3000 Audit Log records.
 Alarm Logs 
The ultrasonic meter stores a log record in the Alarm Log memory whenever an alarm or 
error condition becomes active or is cleared.
For Daniel Mark III Gas Ultrasonic meters, the alarm and error conditions that are logged are 
listed in 
. The ultrasonic meter can store 3000 Alarm Log records.
 System Alarm Logs 
System logs are only available in Daniel Mark III Gas Ultrasonic meters. This log shows 
system related errors such as firmware issues and reasons why writes to the configuration 
may have failed. The meter can store 100 System alarm records.