Emerson Process Management 3422 User Manual

Page of 398
 Reference, Installation, and Operations Manual
Section 7: Maintenance and troubleshooting
3-9000-743 Rev S
June 2013
Installing the transducers
 269
9.
If replacing T-21 or T-22 transducers, install the keyed transformer assembly into the 
transducer holder (
). 
a. Apply a small amount of Molykote 111 to the transformer assembly O-ring.
b.  Insert the keyed transformer into the back end of the transducer holder.   
Figure 7-8  T-22 transducer assembly, holder, transformer, retainer, cable nut and chordset 
10.
For T-21 and T-22 transducers, place the retainer over the transformer assembly. Ensure 
the retainer threads are aligned correctly and hand-tighten. Use a 1 1/8” wrench and 
turn clockwise until the transformer assembly is fully seated in the transducer holder. 
Note: 
Do not over tighten the retainer.
11.
Reconnect the transducer chordset to the transducer holder or the retainer if installing 
T-21 or T-22 transducers. The internal connector of the transducer chordset is keyed 
and will only go on one way. Secure the transducer cabling nut by turning clock-wise. 
Ensure the cable nut threads are correctly aligned.
A.
B.
C.
E.
F.
A. T-21 transformer assy. (W-01 P/N 1-360-03-090) or T-22 transformer assy. (W-02 P/N 1-360-03-110)
B. Transducer holder (Type - H1 P/N 1-360-01-128, H2 P/N 1-360-01-228)
C. Mount and holder
D. Transducer port (meter body)
E. Transducer retainer (Standard P/N 1-360-01-958) High temperature (P/N 1-360-01-978)
D.
F.  Cable nut and chordset:
        100 
o
C - 5’ length P/N 2-3-3400-190)
        100 
o
C -15’ length (P/N 2-3-3400-194
        200 
o
C - Straight backshell (P/N 1-36-01-800)
        200 
o
C - 90
o
 backshell (P/N 1-360-01-801)