Section 7: Maintenance and troubleshooting
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
7.5.3
Chord A to Chord D
Chord(s) A through D - These alarms indicate how a chord is functioning Red and green are the 
only colors used for this alarm.
7.5.4
Field I/O Alarm
Reports various field I/O devices that are in alarm. Click the 
Check Status
 button for more details 
on specific alarms. The field is grayed out if the Daniel 3400 Series Gas Ultrasonic Flow Meter 
does not support this alarm.
7.5.5
Validity Alarm
This alarm indicates that the meter may not be measuring accurately. Click 
Check Status
 to see a 
description of which validity alarms are active. The validity alarms 
QMeter and QFlow indicate 
an issue with the meter collecting enough information from the chords to make an accurate 
measurement. The validity alarms for pressure and temperature indicate that the value is above 
or below the alarm limits for these values. Red and green are the only colors used for this alarm.
7.5.6
Comms Alarm
The Comms alarm indicates that communications between Daniel MeterLink and the meter 
failed. This could be due to a poor communication link. Daniel MeterLink continues to retry 
communications. Red and green are the only colors used for this alarm.
7.5.7
Check Status
Click 
Check Status
 if any of the LEDs are yellow or red to see more specific information causing 
the status alarm. Some alarms do not require an acknowledge and will clear automatically when 
the alarm condition goes away. Alarms that require a user to acknowledge them will have a 
button to the right titled 
ACK
. Clicking the 
ACK 
button changes the button text to 
Wait
 and 
sends a request to the meter to clear the alarm. The alarm will disappear from the Check Status 
dialog once the alarm actually clears.
Table 7-3  Chord alarms
LED Color
Problem
Green
No alarms are present.  Chord is operating properly.
Yellow
At least one sample in the batch caused an alarm but it did not 
cause the chord to fail. The sample will not be used in the batch.  
Discarding occasional samples can occur during normal operation 
such as during flow velocity changes.
Red
The chord has failed or is in acquisition.  This chord is not used for 
this batch. Chords that have failed or are shown to be in acquisi-
tion for repeated batches indicates that the meter should be 
inspected by a service technician.
Gray
The chord has manually been set to inactive.