Emerson Process Management 3422 User Manual

Page of 398
 Reference, Installation, and Operations Manual
Upgrading a Mark II to a Mark III
3-9000-743 Rev S
June 2013
Gain
 307
C.4.3
Gain
The Mark III signal gain values (
GainA1...GainD2) and limits (GainLowLmt and GainHighLmt
are read in hardware gain units. Daniel MeterLink converts the hardware gain units to decibels 
(dB) for display purposes using the conversion equation:
Equation C-10  Mark III Gain Conversion Equation
The Mark III gain units differ from those used by the Mark II. Therefore, the Daniel MeterLink 
configuration conversion does not convert the Mark II 
GainLowLmt
 and 
GainHighLmt
 parameters. 
The Mark III 
GainLowLmt and GainHighLmt data points should be at their default values (the 
widest allowable gain limits) and are unmodified by the Daniel MeterLink configuration 
conversion. These data points can be modified using the Daniel MeterLink 
Edit/Compare 
Configuration
 screen.
C.4.4
Batch update period and stack size
The Mark III batch update occurs at a fixed timed interval whereas the Mark II batch update 
occurs after each 
BatchSize
 firing sequences are completed.
For the Mark III, the batch update period is a function of the 
SpecBatchUpdtPeriod and 
StackSize data points (as described in 
).
When using Daniel MeterLink to convert a Mark II configuration for use in a Mark III, the Mark II 
BatchSize
 parameter is not converted. The Mark III 
SpecBatchUpdtPeriod should be at its 
default value (standard update period of 1 sec) and is unmodified.
The Mark III does not support “stacking-by-chord.” The Mark II 
StackSize
 parameter is converted 
to the Mark III 
StackSize data point as follows:
StackSize
(Mark II)
StackSize
(Mark III)
0, 1, 128 or 129
1 (none)
2 or 130
2
4 or 132
4
8 or 136
8
16, 32, 144, or 160
16
Gain
dB
20
Gain
Hardware
(
)
log
=