Chapter 16
Startup Procedures
16.1
About this chapter
This chapter explains the procedures that are required for initial startup of the Series 3000 device. 
After initial startup, these procedures may be performed as required for troubleshooting or 
maintenance.
The following procedures are discussed:
Applying power to the flowmeter
Sensor zero
Testing the inputs and outputs
Milliamp output trim
Note: If you are using a Model 3300 or Model 3350 controller, the sensor zero procedure is not 
applicable to your device. Perform the other procedures as applicable.
16.2
Applying power
Before you apply power to the Series 3000 device, close and tighten all housing covers.
Turn on the electrical power at the power supply. The flowmeter automatically tests its display. 
During display testing, the screen darkens for approximately five seconds. After the display test is 
completed:
1. The Micro Motion logo is displayed for two to three seconds.
2. An application list is displayed for two to three seconds.
3. The device enters operation mode:
If the discrete batch application is not installed, the process monitor screen is displayed, as 
shown in Figure 17-1. 
If the discrete batch application is installed, the batch process screen is displayed (see 
Figure 18-1).
WARNING
Operating the Series 3000 device without covers in place creates electrical 
hazards that can cause death, injury, or property damage.
Make sure safety barrier partition and covers for the field-wiring, circuit board 
compartments, electronics module, and housing are all in place before applying 
power to the Series 3000 device.