Operation Mode
17.4
Process monitor operation mode
The process monitor, shown in Figure 17-1, is the default operation mode unless the discrete batch 
application is present. 
To scroll through the process monitor, press the left or right cursor control button. The number 
beside each arrow on the top line indicates the screen that will be displayed if the left or right 
cursor control button is pressed. If no display variables are configured for a particular screen, 
that screen is skipped.
If a total appears on the screen, you can press 
RESET
 to reset it to 0. Pressing 
RESET
 will 
reset only the totals on the screen that is displayed.
Press 
PRINT
 to print a ticket that includes values of process variables on the screen that is 
displayed.
You can configure the process variables that will be displayed on each of the five screens of the 
process monitor. To configure the process monitor, see Chapter 12.
If the discrete batch application is installed, you can monitor process variables by using the Process 
Monitor option from the View menu (see Section 17.5).
Figure 17-1
Series 3000 display in process monitor mode
CAUTION
Failure to perform a sensor zero calibration at initial startup could cause the 
flowmeter to produce inaccurate signals.
To avoid inaccurate measurements, perform a sensor zero before putting the 
flowmeter into operation.
Cursor control 
buttons
Security button
Backlit
display
Function buttons
5
DEVICE 1
→2
Mass Flow Rate
2.33
g/s
Mass Total
485.88
g
PRINT
RESET
EXIT
VIEW or EXIT 
button