Emerson Series 3000 MVD User Manual

Page of 348
190
Micro Motion
®
 Series 3000 MVD Transmitters and Controllers
Measurement Performance
21.3.4
Reading and interpreting meter verification test results
Pass/Fail/Abort
When the test is completed, the result will be reported as Pass, Fail, or Abort:
Pass – The test result is within the specification uncertainty limit. In other words, the stiffness 
of the left and right pickoffs match the factory values plus or minus the specification 
uncertainty limit. If transmitter zero and configuration match factory values, the sensor will 
meet factory specifications for flow and density measurement. It is expected that meters will 
pass meter verification every time the test is run.
Fail/Caution – The test result is not within the specification uncertainty limit. Micro Motion 
recommends that you immediately repeat the meter verification test. If you were using Smart 
Meter Verification, with outputs set to Continue Measurement, change the setting to Last 
Measured Value or Fault. 
-
If the meter passes the second test, the first Fail/Caution result can be ignored. 
-
If the meter fails the second test, the flow tubes may be damaged. Use your process 
knowledge to determine the possibilities for damage and the appropriate actions for each. 
These actions might include removing the meter from service and physically inspecting 
the tubes. At minimum, you should perform a flow validation and a density calibration.
Abort – A problem occurred with the meter verification test (e.g., process instability). Abort 
codes are listed and defined in Table 21-3, and suggested actions are provided for each code.
Table 21-3
Meter verification abort codes
Abort code
Description
Suggested action
1
User-initiated abort
None required. Wait for 15 seconds before 
starting another test.
3
Frequency drift
Ensure that temperature, flow, and density are 
stable, and rerun the test.
5
High drive gain
Ensure that flow is stable, minimize entrained 
gas, and rerun the test.
8
Unstable flow
Review the suggestions for stable flow in 
Section 21.3.1 and rerun the test.
13
No factory reference data for meter verification 
test performed on air
Contact Micro Motion customer service and 
provide the abort code.
14
No factory reference data for meter verification 
test performed on water
Contact Micro Motion customer service and 
provide the abort code.
15
No configuration data for meter verification
Contact Micro Motion customer service and 
provide the abort code.
Other
General abort.
Repeat the test. If the test aborts again, contact 
Micro Motion customer service and provide the 
abort code.