Emerson Series 3000 MVD User Manual

Page of 348
222
Micro Motion
®
 Series 3000 MVD Transmitters and Controllers
Diagnostics and Troubleshooting
22.7.4
Configuration alarms
Table 22-8 lists all alarms in the Configuration category, with descriptions, suggested user actions, 
and related information.
Table 22-8
Using configuration alarms
Alarm 
number
Maintenance 
menu listing
Description
Alarm severity
Affected 
by fault 
timeout
User actions
Default
User 
config?
A006
Characterize 
Meter
Master reset has 
been performed. 
Required 
calibration values 
are missing.
Fault
Yes
No
Enter required values. See 
Section 7.3.3.
A020
Cal. Factors 
Missing
Master reset has 
been performed. 
Required 
calibration values 
are missing.
Fault
Yes
No
Enter required values. See 
Section 7.3.3.
A021
Sensor Type 
Incorrect
K1 value missing 
or incorrect, or 
RTD data from 
sensor is 
incorrect.
Fault
No
No
Verify characterization 
parameters. See 
Section 7.3.3.
A027
Security Breach
(Custody transfer 
only) Device has 
been tampered 
with.
Fault
No
No
Weights and Measures 
security seal has been 
broken. Alarm can be 
cleared by user, but 
authorized procedure is 
required to reestablish 
security. Ensure that the 
security switch is set to the 
ON position. Contact Micro 
Motion.
A032
(1)(2)
Meter Verify Fault
Meter verification 
in progress, with 
outputs set to 
fault.
Fault
No
No
Allow the procedure to 
complete. If desired, abort 
the procedure and restart 
with outputs set to last 
measured value.
A032
(3)
Outputs Fixed
Meter verification 
in progress, with 
outputs set to 
Fault or Last 
Measured Value. 
Varies
(4)
Allow the procedure to 
complete.
If desired, abort the 
procedure and restart with 
outputs set to Continue 
Measurement.
A034
Meter Verify 
Failed
The test ran to 
completion but 
the results were 
not within limits.
Info
Yes
No
Rerun the test. If the test fails 
again, see Section 21.3.4.
A035
Meter Verify 
Aborted
The test did not 
run to 
completion.
Info
Yes
No
If desired, read the abort 
code, see Section 21.3.4, 
and perform the appropriate 
action.