Emerson Series 3000 MVD User Manual
222
Micro Motion
®
Series 3000 MVD Transmitters and Controllers
Diagnostics and Troubleshooting
22.7.4
Configuration alarms
Table 22-8 lists all alarms in the Configuration category, with descriptions, suggested user actions,
and related information.
and related information.
Table 22-8
Using configuration alarms
Alarm
number
number
Maintenance
menu listing
menu listing
Description
Alarm severity
Affected
by fault
timeout
by fault
timeout
User actions
Default
User
config?
config?
A006
Characterize
Meter
Meter
Master reset has
been performed.
Required
calibration values
are missing.
been performed.
Required
calibration values
are missing.
Fault
Yes
No
Enter required values. See
Section 7.3.3.
Section 7.3.3.
A020
Cal. Factors
Missing
Missing
Master reset has
been performed.
Required
calibration values
are missing.
been performed.
Required
calibration values
are missing.
Fault
Yes
No
Enter required values. See
Section 7.3.3.
Section 7.3.3.
A021
Sensor Type
Incorrect
Incorrect
K1 value missing
or incorrect, or
RTD data from
sensor is
incorrect.
or incorrect, or
RTD data from
sensor is
incorrect.
Fault
No
No
A027
Security Breach
(Custody transfer
only) Device has
been tampered
with.
only) Device has
been tampered
with.
Fault
No
No
Weights and Measures
security seal has been
broken. Alarm can be
cleared by user, but
authorized procedure is
required to reestablish
security. Ensure that the
security switch is set to the
ON position. Contact Micro
Motion.
security seal has been
broken. Alarm can be
cleared by user, but
authorized procedure is
required to reestablish
security. Ensure that the
security switch is set to the
ON position. Contact Micro
Motion.
A032
(1)(2)
Meter Verify Fault
Meter verification
in progress, with
outputs set to
fault.
in progress, with
outputs set to
fault.
Fault
No
No
Allow the procedure to
complete. If desired, abort
the procedure and restart
with outputs set to last
measured value.
complete. If desired, abort
the procedure and restart
with outputs set to last
measured value.
A032
(3)
Outputs Fixed
Meter verification
in progress, with
outputs set to
Fault or Last
Measured Value.
in progress, with
outputs set to
Fault or Last
Measured Value.
Varies
(4)
Allow the procedure to
complete.
If desired, abort the
procedure and restart with
outputs set to Continue
Measurement.
complete.
If desired, abort the
procedure and restart with
outputs set to Continue
Measurement.
A034
Meter Verify
Failed
Failed
The test ran to
completion but
the results were
not within limits.
completion but
the results were
not within limits.
Info
Yes
No
Rerun the test. If the test fails
again, see Section 21.3.4.
again, see Section 21.3.4.
A035
Meter Verify
Aborted
Aborted
The test did not
run to
completion.
run to
completion.
Info
Yes
No