Installing user-supplied relays
Follow these steps to install user-supplied relays.
Step 1
Connect discrete output terminals to relays
Follow the steps below to connect the transmitter/controller’s discrete output wiring terminals to one, 
two, or three relays.
1. Use the following wire sizes:
Model 3300 or Model 3500: 24 to 16 AWG (0,25 to 1,5 mm
2
)
Model 3350 or Model 3700: 22 to 16 AWG (0,35 to 1,5 mm
2
)
2. Connect the wires between the transmitter/controller and the relay as shown in Table D-2 and 
in the following diagrams:
Model 3300 or Model 3500 with screw-type or solder-tail connectors: see Figure D-8
Model 3300 or Model 3500 with I/O cables: see Figure D-9
Model 3350 or Model 3700: see Figure D-10
WARNING
Shock Hazard.
Improper installation of wiring, or installation with power supply on, can cause 
electric shock or property damage.
For personal and system safety:
Shut off power before installing wiring.
Make sure the installation meets or exceeds local code requirements.
Install relays and wiring in accordance with the illustrations in these instructions.
Install relays and wiring where ambient temperature remains between –4 and 
+140 °F (–20 and +60 °C).
Table D-2
Transmitter/controller terminals for user-supplied relays
Transmitter/controller terminals
Terminal function
Model 3300 or Model 3500
Model 3350 or Model 3700
I/O cables or I/O 
terminal block
Screw-type or 
solder-tail 
terminals
Gray terminal block
19
a 14
20
Return
18
c 14
18
DO1
20
c 16
17
DO2
22
c 18
16
DO3