Configuring Inputs
LD Optimization is a special compensation that is specifically for hydrocarbon liquids. LD 
Optimization should not be used with any other process fluids. LD Optimization is available only with 
certain large sensor sizes. If your sensor can benefit from LD Optimization, the enable/disable option 
will appear in ProLink II or on the display.
7.3.4
Sensor information
Sensor information includes parameters that serve as references without affecting calibration 
parameters, totalizers, or outputs. These parameters are informational and not required.
Sensor information parameters are listed and defined in Table 7-13.
7.3.5
Discrete inputs
Note: Be sure to distinguish these discrete input parameters from the discrete input parameters 
described in Section 7.5, which 
are used to describe the discrete input signal.
The discrete input parameters allow you to specify a transmitter action that will occur when:
An ON signal is received from a discrete input (see Section 7.5 for information on configuring 
the discrete input signal)
A discrete event occurs (see Chapter 10 for information on configuring discrete events)
You can assign one or more actions to a single discrete input or discrete event. All assigned actions 
will be performed. Actions and assignments for discrete inputs and discrete events are listed and 
defined in Table 7-14.
Note: If you are configuring a Model 3300 or Model 3350 controller, refer to Section 7.4 to assign an 
action to a discrete input.
Note: To assign a batch control function to a discrete input or a discrete event, see Chapter 11. To 
assign a print function to a discrete input or discrete event, see Chapter 15.
CAUTION
If you send the transmitter to a calibration facility to perform a water 
calibration, either during startup or any time thereafter, LD Optimization 
must be disabled. When you have completed the calibration, re-enable LD 
Optimization.
Table 7-13
Sensor information parameters
Variable
Default
Description
Sensor model no.
Uninitialized
Automatically set based on calibration data, and cannot be changed.
Sensor serial no.
0
Enter the serial number that is on the sensor serial number tag.
Sensor material
Unknown
Select the appropriate sensor flow tube material.
Sensor flange
Unknown
Select the appropriate flange.
Sensor liner
None
Select the appropriate liner material for the sensor flow tubes.