If you use ProLink II or ProLink III to run a test, a test result chart and a test report are
displayed at the completion of the test. On-screen directions are provided to manipulate
the test data or export the data to a CSV file for offline analysis.
View test result data using the display (standard option)
1.
If you have just run a test, results are displayed automatically at the end of the test.
2.
If you want to view results from previous tests:
a. Navigate to the Smart Meter Verification menu.
Smart Meter Verification – Top-level menu
Figure 9-5:   
Scroll and Select simultaneously 
for 4 seconds
ENTER METER VERFY
Scroll
RUN VERFY
RESULTS READ
SCHEDULE VERFY
Select
EXIT
Scroll
Scroll
Scroll
Scroll
Select
Select
Select
Select
b. Scroll to Results Read and press Select.
The runcount of the most recent test is displayed.
c. To view data for this test, press Select, then press Scroll to scroll through test data.
d. To select a different test, press Scroll, then press Select when the transmitter
displays Results More?. When the desired test appears, as identified by runcount,
press Select.