2.4.2
Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-
character FT value. They are provided on the sensor tag.
character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these may be entered as two
values or as a single 10-character string. The 10-character string is called either Flowcal or
FCF
values or as a single 10-character string. The 10-character string is called either Flowcal or
FCF
.
If your sensor tag shows the FCF and the FT values separately and you need to enter a
single value, concatenate the two values to form the single parameter value.
single value, concatenate the two values to form the single parameter value.
If your sensor tag shows a concatenated Flowcal or FCF value and you need to enter the FCF
and the FT values separately, split the concatenated value:
and the FT values separately, split the concatenated value:
•
FCF
= The first 6 characters, including the decimal point
•
FT
= The last 4 characters, including the decimal point