Tag on older straight-tube sensor (T-Series)
Figure 2-3:   
Tag on newer straight-tube sensor (T-Series)
Figure 2-4:   
2.4.2
Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-
character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these may be entered as two
values or as a single 10-character string. The 10-character string is called either Flowcal or
FCF
.
If your sensor tag shows the FCF and the FT values separately and you need to enter a
single value, concatenate the two values to form the single parameter value.
If your sensor tag shows a concatenated Flowcal or FCF value and you need to enter the FCF
and the FT values separately, split the concatenated value:
FCF
 = The first 6 characters, including the decimal point
FT
 = The last 4 characters, including the decimal point
Quick start
10
 Micro Motion
®