Emerson Model 1700 User Manual

Page of 350
Effect of the Flow Direction parameter and actual flow direction on flow
totals and inventories
Table 4-9:   
Flow Direction
 setting
Actual flow direction
Forward
Zero flow
Reverse
Forward
Totals increase
Totals do not change
Totals do not change
Reverse
Totals do not change
Totals do not change
Totals increase
Bidirectional
Totals increase
Totals do not change
Totals decrease
Absolute Value
Totals increase
Totals do not change
Totals increase
Negate Forward
Totals do not change
Totals do not change
Totals increase
Negate Bidirectional
Totals decrease
Totals do not change
Totals increase
4.5
Configure density measurement
The density measurement parameters control how density is measured and reported.
Density measurement (along with mass measurement) is used to determine liquid volume
flow.
The density measurement parameters include:
Density Measurement Unit
Slug Flow Parameters
Density Damping
Density Cutoff
4.5.1
Configure Density Measurement Unit
Display (standard)
OFF-LINE MAINT
 > OFF-LINE CONFG > UNITS > DENS
Chinese-language
display
Offline Maintain
 > Configuration > Units > Density
ProLink II
ProLink
 > Configuration > Density > Dens Unit
ProLink III
Device Tools
 > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit
 specifies the units of measure that will be displayed for density
measurement.
Procedure
Set Density Measurement Unit to the option you want to use.
Configure process measurement
44
 Micro Motion
®
 Model 1700 Transmitters with Analog Outputs